Model 305/I

Impact Tester Model 305 according to Wegner

Order number: 03050151
Impact tester according to ISO 4532/4586, 5 mm diameter ball, up to 90 N
3190.00 €
Net price plus sales tax. Sale exclusively to commercial buyers

Only small quantities left in stock

Description

The Impact Testing Device – Wegner type, Model 305, is an indispensable aid for testing enamelling in accordance with DIN 51 155.

This practical and handy testing device, which can easily fit into every briefcase, avoids troublesome dismantling and transport. It can be used outside the laboratory or the workshop for testing purposes. Objects can therefore be tested on location and in the position in which they happen to be and where they are in use. This is the only way to test certain parts of enamel surfaces at all, as they are otherwise inaccessible with large and cumbersome testing apparatus.

Execution of Test

In order to set the testing device, the handle is turned until the desired testing force is achieved. This is read on the lower longitudinal scale from 0 - 90 N and the interim values of 1 N on the scale at the girt of the handle. As a comparison, the approached values are shown on the upper scale in Nm, any interim values are to be estimated.

When the spring is tensioned, the impact bolt must be pushed back until it locks into the trigger.

When testing, the device must be held vertically on the surface to be tested with the point resting on the enamel surface. On flat surfaces the point of the device and the two legs will form a triangular setting. If the legs cause problems on uneven surfaces, they can be twisted away or removed, if necessary. To do this the three small knurled screws must be released.

When edges are to be tested, they are laid into the vertically superimposed grooves at the point.

The test objects should be positioned as close as possible to the real situation. Particularly in the case of smaller and light objects care must be taken that they are not clamped tightly.

By pushing the trigger lever, the impact bolt is released for testing.

CAUTION!

With a test force of 30 N and more, the impact bolt should not run ahead without a test object being used, but should always hit an object.

Scope of delivery

  • Impact tester according to Wegner, model 305
  • Plastic case

Standards

1ISO 4532
2DIN ISO 4532
3(DIN 51 155)

Technical data

Drive Manuel
Trigger mechanic
Ball diameter 5 mm
Testing Force 0 - 90 N
Tension device mechanic
Measuring range 0 - 90 N, infinitely variable
Successfully added to watchlist.
Successfully added to cart.