MikroTest 6 Ni 50
Magnetic coating thickness gauge for electroplated nickel layers on non-ferromagnetic substrates, measuring range 0 - 50 μm, automatic measuring process
Coating Thickness Gauge (mechanical)

MikroTest 6 Ni 50

Order number: 05022651
Non-Binding Inquiry
With the button below you can add products to your product inquiry list if you want us to do a quote for this product.
Note this product

Coating Thickness Gauge (mechanical)

MikroTest 6 Ni 50

Magnetic coating thickness gauge MikroTest 6 Ni 50 for fast, accurate and non-destructive measurement of coatings such as:

  • Elelectroplated nickel layers

etc. non-ferromagnetic substrates

  • Measurement without current
  • Calibration-free
  • Can be used in all layers
  • Fully automatic measuring process

Measuring principle and description:

The measurement of the coating thickness is based on a magnetic holding force measurement. The holding force depends on the distance between a permanent magnet and the steel base. This distance is given by the thickness of the non-magnetic layer to be measured.

The magnet is lifted by a measuring spring that exerts a counterforce to the magnetic holding force. The force applied or the angle of rotation of a torsion spring is a direct indicator of the layer thickness. The process is also well applicable for electroplated nickel coatings on copper, aluminium and other non-ferrous metals.

Technical features:

Instrument features that make your quality control even more reliable and accurate:

  • Easy handling
  • High accuracy
  • No calibration necessary. Simply put on and measure
  • Single point measurement
  • No power supply necessary

Technical specifications

Measuring Principle

Magnetic coating thickness measurement

Layer Thickness

0 ... 50 μm

Material

Electroplated nickel layers on non-ferromagnetic substrates