DAkkS accredited calibration acc. to ISO 376:2011 Force range: up to 25 kN (with display)
DAkkS accredited calibration - force
DAkkS accredited calibration one load direction with 10 force levels
Order number: 16310232-
Non-Binding Inquiry
With the button below you can add products to your product inquiry list if you want us to do a quote for this product.
Note this product
DAkkS accredited calibration - force
DAkkS accredited calibration one load direction with 10 force levels
DAkkS accredited calibration acc. to ISO 376:2011 Force range: up to 25 kN (with display).
- Smallest measurement uncertainty < 5 N: 0,15%
- Smallest measurement uncertainty ≥ 5 N: 0,075%
- Smallest measurement uncertainty ≥ 200 N: 0,1%
The calibration certificate contains a graphical representation of the calibration results (mean value incl. uncertainty interval as well as specification limit).
If a different or no representation is desired, please specify when ordering. If a class division is desired, please specify this when ordering.